Health product recall

Quart Arterial Filter (2019-08-29)

Starting date:
August 29, 2019
Posting date:
September 6, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70913


Last updated:

Affected Products

Quart Arterial Filter

Reason

Maquet cardiopulmonary has determined that the sterile barrier system of the Quart Arterial Filter may be compromised during transportation. During verification testing of transportation simulation, the integrity of the sterile packaging of one sample of 48 tested resulted in a damaged sterile barrier system of the product. Exposure to a non-sterile or potentially non-sterile medical device may result in infection-causing inflammatory like syndromes thereby deteriorating the clinical state of the patient. Additionally, infection may occur if the device is connected to the central circulatory system. Individuals undergoing extracorporeal circulation usually develop inflammatory response due to the fact that human blood cells are exposed to foreign surface with a release of inflammatory mediators as the consequence. The most severe form is called systemic inflammatory response syndrome (SIRS).  Maquet Cardiopulmonary has not received any complaints associated with damage to the sterile barrier system or to serious injuries or death due to damage to the sterile barrier.

Affected products

Quart Arterial Filter

Lot or serial number

70119626

70126255

70126256

70126309

Model or catalog number

HBF 140

Companies
Manufacturer

Maquet Cardiopulmonary GmbH

Kehler Str. 31

Rastatt

76437

GERMANY