This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

QUANTA LYSER SOFTWARE (2015-07-15)

Starting date:
July 15, 2015
Posting date:
August 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54548

Affected Products

A. QUANTA LYSER SOFTWARE

Reason

As a result of a complaint and subsequent investigation, a software defect was detected that affects all versions of QUANTA-LYSER software when processing EIA Assays in duplicate only. Results reported from the QUANTA-LYSER using QUANTA link data management software are not affected.

Affected products

A. QUANTA LYSER SOFTWARE

Lot or serial number

VERSION 3.2 AND 3.1

Model or catalog number
  • 066QL0171
  • 066QL0175
Companies
Manufacturer

Inova Diagnostics Inc.

9900 Old Grove Road

San Diego

92131

California

UNITED STATES