Health product recall

QUANTA Flash DFS70 Controls (2018-11-10)

Starting date:
November 10, 2018
Posting date:
March 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69308

Affected Products

QUANTA Flash DFS70 Controls

Reason

Upon investigation into complaints from customers, Inova Diagnostics has confirmed that the positive control value on lots 171589, 181676 and 181799 of QUANTA Flash DFS70 Controls fall below the acceptable range. When control values fall outside of the acceptable range, the run is considered invalid. Results from an invalid run cannot be reported and the run needs to be repeated.

Affected products

QUANTA Flash DFS70 Controls

Lot or serial number
  • 171589
  • 181676
  • 181799
Model or catalog number

701267