Health product recall

QUANTA FLASH DFS70 (2016-04-23)

Starting date:
April 23, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58366

Affected products 

  1. QUANTA FLASH DFS70

Reason

Control values for lot 151011 of QUANTA Flash DFS70 fell below the acceptable range. When control values fall outside of the acceptable range, the run is considered invalid. Results from an invalid run cannot be reported and the run needs to be repeated.

Affected products

  1. QUANTA FLASH DFS70

Lot or serial number
  • 151011
Model or catalog number
  • 701268
Companies
Manufacturer

Inova Diagnostics Inc.

9900 Old Grove Road

San Diego

92131

California

UNITED STATES