Health product recall

QAngio XA 3D

Last updated

Summary

Product
QAngio XA 3D
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
QAngio XA 3D Not applicable 2.2.2.4

Issue

For certain Siemens XA images, the isocenter calibration value is calculated incorrectly due to an incorrect value for distance source to isocenter that is saved by Siemens in the standard dicom attribute (0018,1111). The correct value is stored in a Siemens private dicom attribute (0021, xx17, "Siemens sms-ax acq 1.0") in the dicom data. Incorrect calibration can lead to errors in the measurements provided by qAngio XA 3D.

Recall start date: December 16, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Medis Medical Imaging Systems B.V.
Schuttersveld 9, Leiden, Netherlands, 2316 XG
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76759