Health product recall

Q-Rad System (2019-03-30)

Starting date:
March 30, 2019
Posting date:
April 5, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69476


Last updated:

Affected Products

Q-Rad System

Reason

Carestream Health has discovered a potential safety problem that could lead to the breakage of the two steel cables that attach the equipment to its ceiling support mechanism. It has been determined that in some instances when the main steel rope breaks, the safety lock may not be triggered and the safety rope is then providing the only support for the lifting column and support arm. This safety rope is not designed for continuous load and on-going movement of the support arm leading to potential serious injury for users and patients, should the safety rope also fail.

Affected products

Q-Rad System

Lot or serial number
  • RS590LD SN Range RS590LD12K-1002 to RS590LD14N-1005.
  • RS580 SN Range 1039 to 5149.
Model or catalog number

More than 100 numbers, contact manufacturer.

Companies
Manufacturer

Carestream Health, Inc.

150 Verona Street

Rochester

14608

New York

UNITED STATES