Provido, Proveo 8 and M530 OHX Microscopes
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
M530 OHX – Optics Carrier |
10448737 |
291122001 |
Leica Proveo 8 Ophthalmic Microscope System |
10449058 |
More than 10 numbers, contact manufacturer. |
Provido |
10448976 |
110722001 |
Issue
Complaint trending revealed an adverse trend of complaints stating that a LMS surgical operating microscope unexpectedly shut down caused by a defective 24VDC power supply which is used inside the surgical operating microscopes Provido, Proveo 8 and M530 OHX. Based on the results of the investigations there is a reasonable probability that the MOSFET in the power supply unit of these microscopes will overheat and become defective.
Recall Start Date: August 30, 2023
Additional information
Details
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201, Heerbrugg, St. Gallen, Switzerland, 9435
Get notified
Receive emails about new and updated recall and safety alerts.