This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Protektor IOM and Protektor 32 System (2014-10-29)

Starting date:
October 29, 2014
Posting date:
November 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42709

Recalled Products

A) Protektor IOM system

B) Protektor 32 System

Reason

When using remote monitoring with local acquisition system configured to block remote acquisition changes,  if the remote user tries to stop the free run waveform group, the system will display a message informing the user that they do not have sufficient privilege to perform this action.

Affected products

A) Protektor IOM system

Lot or serial number

PRACQ0103C8-S

PRACQ0108C9-S

PRACQ0120D3-S

PRACQ0438E

Model or catalog number

10271

Companies
Manufacturer

Natus Medical Incorporated DBA Excel-Tech Ltd (XLTEK),

2568 Bristol Circle,

Oakville,

L6H 5S1

Ontario

CANADA



B) Protektor 32 System

Lot or serial number

00298000102B

0029800157B

0029800164B

0029800176B

002980B20291

002980B20292

002980B20355

Model or catalog number

002980

Companies
Manufacturer

Natus Medical Incorporated DBA Excel-Tech Ltd (XLTEK),

2568 Bristol Circle,

Oakville,

L6H 5S1

Ontario

CANADA