Health product recall

Primus Anaesthetic Workstation

Last updated

Summary

Product
Primus Anaesthetic Workstation
Issue
Medical devices - Manufacturing defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
Primus Infinity Empowered (Anesthesia Workstation) ASMH-0113 8607500
Scio Four Oxi Plus More than 10 numbers, contact manufacturer. 68 71 801
Primus Anaesthetic Workstation ASLK-0157 8603800-XX
Vamos/ Vamos Plus ASNL-6553, ASNN-6506, ASNN-6507, ASPB-6571, ASNL-6552, ASNL-6551, ASPB-6539 6873350

Issue

During global post market surveillance activities cases have been reported that the sampling pump 6873493 shows a changing characteristic with increasing operating time, which can lead to the deactivation of the gas measurement during operation. The manufacturer has released a new version of pump and the new pump will be replaced on all affected devices.

Recall start date: Apr 21, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Health products - Medical devices - Cardiovascular
Companies

Draegerwerk Ag & Co. Kgaa

Moislinger Allee 53-55, Lubeck

Germany, 23542

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64157

Get notified

Receive notifications for new and updated recalls and alerts by category.