Primus Anaesthetic Workstation
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
||
---|---|---|---|---|
Primus Infinity Empowered (Anesthesia Workstation) |
ASMH-0113 |
8607500 |
||
Scio Four Oxi Plus |
More than 10 numbers, contact manufacturer. |
68 71 801 |
||
Primus Anaesthetic Workstation |
ASLK-0157 |
8603800-XX |
||
Vamos/ Vamos Plus |
ASNL-6553, ASNN-6506, ASNN-6507, ASPB-6571, ASNL-6552, ASNL-6551, ASPB-6539 |
6873350 |
Issue
During global post market surveillance activities cases have been reported that the sampling pump 6873493 shows a changing characteristic with increasing operating time, which can lead to the deactivation of the gas measurement during operation. The manufacturer has released a new version of pump and the new pump will be replaced on all affected devices.
Recall start date: Apr 21, 2022
Additional information
Details
Draegerwerk Ag & Co. Kgaa
Moislinger Allee 53-55, Lubeck
Germany, 23542