Health product recall

Primus Anaesthetic Workstation

Last updated

Summary

Product
Primus Anaesthetic Workstation
Issue
Medical devices - Manufacturing defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

Primus Infinity Empowered (Anesthesia Workstation)

ASMH-0113

8607500

Scio Four Oxi Plus

More than 10 numbers, contact manufacturer.

68 71 801

Primus Anaesthetic Workstation

ASLK-0157

8603800-XX

Vamos/ Vamos Plus

ASNL-6553, ASNN-6506, ASNN-6507, ASPB-6571, ASNL-6552, ASNL-6551, ASPB-6539

6873350

Issue

During global post market surveillance activities cases have been reported that the sampling pump 6873493 shows a changing characteristic with increasing operating time, which can lead to the deactivation of the gas measurement during operation. The manufacturer has released a new version of pump and the new pump will be replaced on all affected devices.