Primus Anaesthetic Workstation
Brand(s)
Last updated
Summary
Product
Primus Anaesthetic Workstation
Issue
Medical devices - Manufacturing defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products | Lot or Serial Number | Model or catalog number | ||
---|---|---|---|---|
Primus Infinity Empowered (Anesthesia Workstation) | ASMH-0113 | 8607500 | ||
Scio Four Oxi Plus | More than 10 numbers, contact manufacturer. | 68 71 801 | ||
Primus Anaesthetic Workstation | ASLK-0157 | 8603800-XX | ||
Vamos/ Vamos Plus | ASNL-6553, ASNN-6506, ASNN-6507, ASPB-6571, ASNL-6552, ASNL-6551, ASPB-6539 | 6873350 |
Issue
During global post market surveillance activities cases have been reported that the sampling pump 6873493 shows a changing characteristic with increasing operating time, which can lead to the deactivation of the gas measurement during operation. The manufacturer has released a new version of pump and the new pump will be replaced on all affected devices.
Recall start date: Apr 21, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Health products - Medical devices - Cardiovascular
Companies
Draegerwerk Ag & Co. Kgaa
Moislinger Allee 53-55, Lubeck
Germany, 23542
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64157