Health product recall

Previ Isola (2014-01-31)

Starting date:
January 31, 2014
Posting date:
March 19, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38473

Recalled Products

Previ Isola

Reason

It has been determined that the pipettor may fail in very few cases (Less than 1%) to dispense a urine sample due to a sample ''wicking'' where the dispensed sample remains adhered to the side of the pipette tip rather than dropping to the agar surface. In such cases, the first plate may have no sample dispensed and a double volume may be dispensed on the next plate if a multiple plate panel is being used.

Affected products

Previ Isola

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

AS180-00001 TO AS180-00476

Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

84894

Missouri

UNITED STATES