Health product recall

PLUM 360 INFUSION SYSTEM WITH MEDNET (2017-10-19)

Starting date:
October 19, 2017
Posting date:
November 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, Hospitals, General Public
Identification number:
RA-65066

Affected products

PLUM 360 INFUSION SYSTEM WITH MEDNET

Reason

Issue 1: The user is not able to turn off the pump when a pump in a distal occlusion paused state is exposed to a second "power off then on" malfunction.

Issue 2: When the user accesses the preventive maintenance screen, in service/biomed mode, on Plum 360 v15.10.00.010 devices, with a total delivery time greater than 1,500 hours, the user will not be able to interact with the device and the device must be power cycled.

Affected products

PLUM 360 INFUSION SYSTEM WITH MEDNET

Lot or serial number

SOFTWARE V15.10.00.010

Model or catalog number

30010-XX-YY

Companies
Manufacturer

ICU Medical Inc.

600 NORTH FIELD DRIVE

LAKE FOREST

60045

Illinois

UNITED STATES