Health product recall

Plum 360 Infusion System

Starting date:
June 6, 2017
Posting date:
July 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63856

Reason

ICU Medical has identified a potential for the connectivity engine (CE) module to disengage from the main chassis. In a rare situation, this could lead the Plum 360 Infuser to power down without an alarm notification resulting in a delay of either initiation of a therapy or interruption of an active infusion.

Affected products

Plum 360 Infusion System

Lot or serial number

All Serial numbers

Model or catalog number

30010

Companies

Manufacturer
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES