This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Plum 360 Infusion System

Starting date:
June 6, 2017
Posting date:
July 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63856

Reason

ICU Medical has identified a potential for the connectivity engine (CE) module to disengage from the main chassis. In a rare situation, this could lead the Plum 360 Infuser to power down without an alarm notification resulting in a delay of either initiation of a therapy or interruption of an active infusion.

Affected products

Plum 360 Infusion System

Lot or serial number

All Serial numbers

Model or catalog number

30010

Companies

Manufacturer

Hospira Inc.

275 North Field Drive

Lake Forest

60045

Illinois

UNITED STATES