Plum 360 Infusion System
- Starting date:
- June 6, 2017
- Posting date:
- July 7, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63856
Reason
ICU Medical has identified a potential for the connectivity engine (CE) module to disengage from the main chassis. In a rare situation, this could lead the Plum 360 Infuser to power down without an alarm notification resulting in a delay of either initiation of a therapy or interruption of an active infusion.
Affected products
Plum 360 Infusion System
Lot or serial number
All Serial numbers
Model or catalog number
30010
Companies
- Manufacturer
-
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES