Health product recall

Platinum (DRF) (2014-12-11)

Starting date:
December 11, 2014
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52573

Recalled products

  1. Platinum (DRF)

Reason

The X-Ray Tube / Collimator became dissociated from its support, causing it to fall out of position. The assembly was retained by the cables and there was no collision with the patient's support panel. The probable cause for the failure is the fragility of the welding of part M009G001 "Axe Support Tube".

Affected products

A. Platinum (DRF)

Lot or serial number
  • C116F007
Model or catalog number
  • 6F
Companies
Manufacturer
Apelem SAS,
Parc Scientifique Georges Besse, 175,
Allée Von Neumann, F30035, Nîmes
30900
FRANCE