This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Platinum (DRF) (2014-12-11)

Starting date:
December 11, 2014
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52573

Recalled products

  1. Platinum (DRF)

Reason

The X-Ray Tube / Collimator became dissociated from its support, causing it to fall out of position. The assembly was retained by the cables and there was no collision with the patient's support panel. The probable cause for the failure is the fragility of the welding of part M009G001 "Axe Support Tube".

Affected products

A. Platinum (DRF)

Lot or serial number
  • C116F007
Model or catalog number
  • 6F
Companies
Manufacturer

Apelem SAS,

Parc Scientifique Georges Besse, 175,

Allée Von Neumann, F30035, Nîmes

30900

FRANCE