Platelia Aspergillus IgG
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Platelia Aspergillus IgG |
2L0039 |
62783 |
Issue
Bio-Rad Laboratories is initiating a recall due to a confirmed quality anomaly for Platelia aspergillus IgG test (cat #62783), for lot 2l0039 (expiry date: 2024/04/15). The current lot 2l0039 shows a decrease of Optical Density (OD) of all calibrators. The OD of the R4A calibrator 10 au/ml could be lower than 0.200 (the expected quality criteria is R4A OD value > 0.200) that could result in an invalid run.
Recall start date: May 11, 2023