This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

PK7300 Automated Microplate System (2017-06-19)

Starting date:
June 19, 2017
Posting date:
July 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63864

Affected Products

A. PK7300 Automated Microplate System

B. Beckman Coulter PK7300 Automated Microplate System

Reason

Beckman Coulter has confirmed that, during sample processing under certain conditions, the PK7300 does not always generate a reagent dispensing pressure error when the reagent vial becomes empty and no reagent is dispensed.

Sample processing will continue after a vial is empty only when the reagent alarm volume, limit volume and associated mode shift alarm have not been set.

Affected products

A. PK7300 Automated Microplate System

Lot or serial number

0030221

0060227

0070235

0070236

0110255

1010262

2013030365

9010177

9110205

Model or catalog number

PK7300

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



B. Beckman Coulter PK7300 Automated Microplate System

Lot or serial number

0030221

0060227

0070235

0070236

0110255

1010262

2013030365

9010177

9110205

Model or catalog number

PK7300

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES