Phoroptor VRx Digital Refraction System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Phoroptor VRx Digital Refraction System |
78850-1221 |
16242 |
Issue
Some Phoroptor VRx devices may have an assembly defect that causes the device to detach from the bracket that connects it to an ophthalmic stand. A device that detaches from a stand can fall on to a patient or operator causing an injury. The defect was identified via a customer complaint received on April 3, 2023. The subsequent complaint investigation showed that other units built in December 2021 and January 2022 could have the same defect.
Recall Start Date: August 11, 2023
Additional information
Details
Reichert, Inc.
3362 Walden Ave., Depew, New York, United States, 14043
Get notified