Health product recall

Philips BiPAP and Humidifiers

Last updated

Summary

Product
Philips BiPAP and Humidifiers
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
BiPAP A40 Pro, CA All lots. CAX3100S12
BiPAP A40, Canada, Core Package All lots. 1111174
BiPAP A30 System - Ventilator & System One Aseries Heated Humidifier All lots. 1111148
BiPAP A40 EFL All lots. INX3T
System One Heated Humidifier All lots. IN6T INX3T IN6TFLG IN6H INX3H IN6HFLG
BiPAP A40 Pro, CA W/ Heated Tube Humidifier, CA All Lots. CAX3100T12
System One A-Series Heated Humidifier All Lots. 1111552 1077107 1111555

Issue

There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device components are unknown. Full validation requires biological evaluation after cleaning cycles per ISO 10993-1 and testing on expected worse-case clinical use per ISO 18562-1. the maximum number of Keredusy cycles was also not identified within the user manuals. ISO 17664-1 does allow for the manufacturer's IFU to direct the processor to refer to the disinfecting agent manufacturer's IFU with reference to concentration, temperature, and contact time, but these parameters were not defined nor documented within the IFU. The IFU only directed the user to Philips customer service.

Recall Start Date: June 30, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Respironics Inc. Also Trading As Philips Rs North America LLC

1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73914