PHEON MODULAR KNEE JOINTS (2015-07-24)

Starting date:
July 24, 2015
Posting date:
August 11, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54588

Affected Products

  1. PHEON MODULAR KNEE JOINT 3R62
  2. PHEON MODULAR KNEE JOINT 3R62=1

Reason

Issue identified where locking function may fail if it is activated if the locking pin is not completely pulled out by patient; edge loading in locking surface causes a deformation of locking pin. The pin may get stuck in the guide of the upper section hindering the lock from engaging automatically in the extended position. The stability of the joint is then only by alignment and polycentric joint design and may result in buckling of the loaded joint.

Affected products

A. PHEON MODULAR KNEE JOINT 3R62

Lot or serial number

N/A

Model or catalog number

3R62

Companies
Manufacturer

Otto Bock Healthcare GMBH

MAX NADER STRABE 15

DUDERSTADT

37115

GERMANY



B. PHEON MODULAR KNEE JOINT 3R62=1

Lot or serial number

N/A

Model or catalog number

3R62=1

Companies
Manufacturer

Otto Bock Healthcare GMBH

MAX NADER STRABE 15

DUDERSTADT

37115