Reason

Customer complaints have reported a decrease of values for EliA ANA positive control when using EliA dsDNA well lot bfa3y/0142. An internal investigation showed that EliA dsDNA well lot bfa3y/0142 gives higher results, and in some cases the results may be 3 times higher than other dsDNA well lots. There is potential for a false positive test result.