Peritoneal Dialysis(PD) Swan Neck Catheters (2014-11-10)
- Starting date:
- November 10, 2014
- Posting date:
- December 5, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42771
Recalled Products
Peritoneal Dialysis(PD) Swan Neck Catheters
Reason
Medionics has received a few reports of opacification and ballooning of the peritoneal dialysis catheters at the exit-site, however no adverse events have been reported associated with this issue. The age of the implanted catheters in the reported incidents ranged from 30 months to 80 months.
Investigation of these incidents indicates that the structural changes to PD catheters at the exit-site may occur with prolonged use of certain cleansing agents and antibiotic cream or ointment. Additionally, use of solutions containing organic solvents for catheter cleaning may cause the breakdown of silicone tubing. The likelihood of structural changes to silicone PD catheters increases over the duration of implantation.
Affected products
Peritoneal Dialysis(PD) Swan Neck Catheters
Lot or serial number
All lots,contact manufacturer.
Model or catalog number
SNCA5262L
Companies
- Manufacturer
-
Medionics International Inc.,
114 Anderson Avenue,
Markham
L6E 1A5
Ontario
CANADA