This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

PERCEVAL SUTURELESS HEART VALVE (2016-08-12)

Starting date:
August 12, 2016
Posting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59986

Affected Products

PERCEVAL SUTURELESS HEART VALVE

Reason

Livanova Canada Corp. (lnc) has identified a manufacturing issue affecting a limited number of perceval valves. These valves were shipped in a sterile packing solution that may contain higher than expected residual levels of homocysteic acid. The homocysteic acid is used during the manufacturing process to neutralize the free aldehydes introduced by the sterilization process.

Affected products

PERCEVAL SUTURELESS HEART VALVE

Lot or serial number
  • A29638
  • A30436
  • A31487
  • A31846
Model or catalog number
  • PVS21
  • PVS23
Companies
Manufacturer

LIVANOVA CANADA CORP.

5005 NORTH FRASER WAY

BURNABY

British Columbia

CANADA