This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Pentax Video Duodenoscope

Starting date:
March 16, 2017
Posting date:
April 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63012

Reason

During interactive discussions with US FDA concerning a 510(k) additional information ("AI") request for the ED-3490TK duodenoscopes, the FDA requested that Pentax perform a root cause analysis to determine how soiling occurred on the surface of the suction cylinder and under the distal cap during testing. FDA also requested that Pentax identify risk mitigation strategies to address the manufacturing process, possible damage to the duodenoscope with use, and any other aspects of risk mitigation. As a result of the communications with FDA, and at the suggestion of FDA, it was decided that Pentax would issue a customer safety communication to advise customers who use ED-3490TKS about a potential issue associated with the distal cap and offer recommendations intended to reduce the potential risk for contamination and subsequent patient infection. Specifically, the user communication states, "during manufacturing of ED-3490TK duodenoscopes, silicone adhesive is applied to the distal tip prior to affixing the distal cap. In some instances, cracks or gaps may form in the adhesive that may be vulnerable to fluid ingress and soiling." the letter reminds users of the importance of using the ed-3490tk in accordance with its current labeling and assuring that all reprocessing personnel are knowledgeable and thoroughly trained on the current IFU for manual reprocessing of these devices. Additionally, the letter recommends that users "immediately remove from use any ed-3490tk duodenoscope that shows visible signs of wear or physical damage" and states that Pentax will provide on-site inspections to assess duodenoscope functionality, with particular attention to the integrity of the distal cap, and provide in-house servicing as needed.

Affected products

Pentax Video Duodenoscope

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

ED-3490TK

Companies

Manufacturer

Hoya Corporation

6-10-1 NISHI-SHINJUKU

Shinjuku-ku

160-0023

JAPAN