This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Pentax Medical Endoscopes (2017-07-17)

Starting date:
July 17, 2017
Posting date:
August 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64158

Affected Products

  1. Slim Cystoscope Set
  2. Pentax Vb-2000 Video Bronchoscope
  3. E.N.T. Scope

Reason

Pentax Medical has become aware that some customers may own legacy endoscopes where the screw connecting the suction arm to the control body could loosen with time. A loose suction arm may result in inadequate suctioning due to leakage of air. There is also the potential for organic debris to accumulate in the space between the suction nipple and control body. In some cases, these events could cause cross-contamination between patients.

Affected products

  1. Slim Cystoscope Set

Lot or serial number
  • G110026
  • G110030
  • G110031
  • G110036
  • G110041
  • G110042
  • G110046
  • G110050
Model or catalog number

ECY-1570K

Companies
Manufacturer

Hoya Corporation

6-10-1 NISHI-SHINJUKU

Shinjuku-ku

160-0023

JAPAN



B. Pentax Vb-2000 Video Bronchoscope

Lot or serial number

G110017

G110029

G110123

G110458

G110589

G110907

G111020

G111021

Model or catalog number

EB-117OK

EB-1570K

EB-1970K

EB-1970TK

Companies
Manufacturer

Hoya Corporation

6-10-1 NISHI-SHINJUKU

Shinjuku-ku

160-0023

JAPAN



C. E.N.T. Scope

Lot or serial number

G110059

G110117

Model or catalog number

VNL-1570STK

Companies
Manufacturer

Hoya Corporation

6-10-1 NISHI-SHINJUKU

Shinjuku-ku

160-0023

JAPAN