PD-L1 IHC 22C3 pharmDx
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected product |
Lot or serial number |
Model or catalogue number |
|---|---|---|
PD-L1 IHC 22C3 pharmDx |
12001736B |
GE006 |
PD-L1 IHC 22C3 pharmDx |
12010368B |
GE006 |
PD-L1 IHC 22C3 pharmDx |
11992577B |
GE006 |
Issue
Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots.