Pas Staining Kit
Brand(s)
Last updated
Summary
Product
Pas Staining Kit
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Pas Staining Kit |
More than 10 numbers, contact manufacturer. |
860-014 |
Issue
Internal stability testing was performed with the Pas Staining Kit where it was confirmed that the stability of 24 months could not be met. The expiration date of the product was reduced to 12 months after the manufacturing date.
Recall start date: June 22, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Ventana Medical Systems Inc
1910 E Innovation Park Drive, Tucson, Arizona, United States, 85755
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-64373