This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Paradigm Insulin Pumps and Paradigm Insulin Infusion Pump (2014-03-17)

Starting date:
March 17, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39013

Recalled Products

A. Paradigm 522 Insulin Pump

B. Paradigm 722 Insulin Pump

C. Paradigm Insulin Infusion Pump

D. Paradigm 554 Insulin Pump

E. Paradigm 754 Insulin Pump

Reason

Purpose of this notification is to reinforce important information regarding the safe use of Medtronic insulin pumps. Medtronic has received a small number of reports regarding users who have unintentionally programmed the pump to deliver the maximum bolus amount, including one incident that resulted in severe hypoglycemia. The inadvertent selection and delivery of a bolus amount that exceeds the amount intended by the user can result in hypoglycemia.

Medtronic is notifying all Paradigm pump users Healthcare professionals and certified pump trainers of the function of the wrap around scrolling feature and make them aware of the potential for user error.

Affected products

A. Paradigm 522 Insulin Pump

 

Lot or serial number

All lots

Model or catalog number

MMT-522

 

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES



B. Paradigm 722 Insulin Pump

 

Lot or serial number

All lots

Model or catalog number

MMT-722

 

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES



C. Paradigm Insulin Infusion Pump

 

Lot or serial number

All lots

Model or catalog number

MMT-511

MMT-512

MMT-515

MMT-712

MMT-715

 

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES



D. Paradigm 554 Insulin Pump

 

Lot or serial number

All lots

Model or catalog number

MMT-554

 

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES



E. Paradigm 754 Insulin Pump

Lot or serial number

All lots

Model or catalog number

MMT-754

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES