Paradigm Insulin Infusion Pumps Remote Programmer (2018-08-13)
- Starting date:
- August 13, 2018
- Posting date:
- September 7, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67684
Affected products
Paradigm Insulin Infusion Pumps Remote Programmer
Reason
An external security researcher has identified a potential vulnerability related to the MiniMed Paradigm Family of insulin pumps and corresponding remote controller. The researcher's report states that an unauthorized individual in close proximity of an insulin pump user could potentially copy the wireless Radio Frequency (RF) signals from the user's remote controller (while they are in the process of delivering a remote bolus) and play those back later to deliver an involuntary bolus of insulin to the user. This could lead to potential health risks such as hypoglycemia if additional insulin is delivered beyond the user's insulin requirements.
Affected products
Paradigm Insulin Infusion Pumps Remote Programmer
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- MMT-503
- MMT-503NA
Companies
- Manufacturer
-
MEDTRONIC MINIMED