Health product recall

Panther System

Brand(s)
Last updated

Summary

Product
Panther System
Issue
Medical devices - Performance issue
What to do

Company has notified customer of this situation, and allowed for software update. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Panther System

Not applicable.

303095

Issue

According to the assay package inserts for the following Aptima assays: HIV-1 Quant DX, HBV Quant, HCV Quant DX, CMV Quant, BV, and CV/TV, controls from different lot numbers can be used. However, Hologic has identified a software anomaly with Panther/Fusion SW  version 7.2.9, where assay processing will not start when the control lots loaded on the Panther/Panther Fusion system do not match the ones listed on the control barcode sheet.

Recall start date: November 17, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Hologic, Inc.
10210 Genetic Center Drive, San Diego, California, United States, 92121
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75066