Paklx - MATCH IT! Platelet Antibody Software (V1.1) (2017-09-29)
- Starting date:
- September 29, 2017
- Posting date:
- January 11, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Natural health products
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65560
Affected Products
Paklx - MATCH IT! Platelet Antibody Software (V1.1)
Reason
The Paklx Instructions For Use (IFU), version 303285.IFUEN REV K, and the MATCH IT! Platelet Antibody Software database were not aligned. The IFU states that a minimum of 60 events must be collected per bead region. The MATCH IT! Platelet Antibody Software database requires a minimum of 100 events per bead region. This discrepancy caused invalid results when using the MATCH IT! Platelet antibody software to analyze a bead.
Affected products
Paklx - MATCH IT! Platelet Antibody Software (V1.1)
Lot or serial number
3004895
3005391
3005538
Model or catalog number
303285
888622
PLX
Companies
- Manufacturer
-
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186