Health product recall

Pagewriter Electrocardiograph

Last updated

Summary

Product
Pagewriter Electrocardiograph
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or Serial number

Model or Catalogue number

Pagewriter Tc70 Electrocardiograph

Not applicable.

PAGEWRITER TC70

Pagewriter Tc50 - Electrocardiograph

Not applicable.

PAGEWRITER TC50

Issue

The pagewriter tc30/50/70 cardiograph device could potentially fail to charge while plugged into the ac power outlet. This is due to a faulty mainboard. based on a supplier analysis, the faulty mainboard was caused by delamination of an ic chip after mold assembly.

Recall start date: May 19, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Philips Medical Systems

3000 Minuteman Road, Andover, Massachusetts, Unites States, 01810

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type III
Identification number
RA-64204

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe