Health product recall

OXYLOG 3000 SYSTEM - DISPOSABLE VENTILATION HOSE (2016-06-02)

Starting date:
June 2, 2016
Posting date:
June 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-58926

Affected products

  1. OXYLOG 3000 SYSTEM - DISPOSABLE VENTILATION HOSE

Reason

During the course of routine internal testing, the firm detected isolated episodes of leakage at the check valve (one way valve of the inspiration branch) of the disposable pediatric ventilation hose of Oxylog 3000 system and Oxylog plus, model 5704964. Affected circuits with this fault were potentially distributed to customers. All other ventilation circuits for the Oxylog family of devices are not affected.

Leakage at the check valve can result in patient's exhaled gas entering into the breathing circuit, which could lead to the rebreathing of the same exhaled gas with reduced oxygen concentration/elevated CO2 for the patient. This leakage is not detected during the ventilator operational readiness check.