OXYLOG 3000 SYSTEM - DISPOSABLE VENTILATION HOSE (2016-06-02)
- Starting date:
- June 2, 2016
- Posting date:
- June 16, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-58926
Affected products
- OXYLOG 3000 SYSTEM - DISPOSABLE VENTILATION HOSE
Reason
During the course of routine internal testing, the firm detected isolated episodes of leakage at the check valve (one way valve of the inspiration branch) of the disposable pediatric ventilation hose of Oxylog 3000 system and Oxylog plus, model 5704964. Affected circuits with this fault were potentially distributed to customers. All other ventilation circuits for the Oxylog family of devices are not affected.
Leakage at the check valve can result in patient's exhaled gas entering into the breathing circuit, which could lead to the rebreathing of the same exhaled gas with reduced oxygen concentration/elevated CO2 for the patient. This leakage is not detected during the ventilator operational readiness check.
Affected products
A. OXYLOG 3000 SYSTEM - DISPOSABLE VENTILATION HOSE
Lot or serial number
- MFG DATES
Model or catalog number
- 5704964
Companies
- Manufacturer
-
DRAEGERWERK AG & CO
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY