Health product recall

ORT (2021-09-20)



Starting date:
September 20, 2021
Posting date:
October 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76559


Last updated:

Affected Products

ORT300

Reason

The ORT200/300 operating room table is specifically designed to provide an mr compatible platform for neurosurgical use for cranial and spinal surgeries. The tables are removable from the operating room and mate to a docking station installed in the floor of the or via a floor latch mechanism which contains 4 linear actuators. Complaints have reported an inability to dock or undock the operating room table from the floor baseplate, due to malfunction of the actuators within the floor latch locking mechanism.

Affected products

ORT300

Lot or serial number

Not applicable.

Model or catalog number

114093-600

Companies
Manufacturer

Deerfield Imaging, Inc.

5101 Shady Oak Road

Minnetonka

55343

Minnesota

UNITED STATES