OpTmizer CTS T-Cell Expansion SFM (2019-05-15)
- Starting date:
- May 15, 2019
- Posting date:
- May 31, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70071
Last updated: 2019-05-31
Affected Products
OpTmizer CTS T-Cell Expansion SFM
Reason
Defective machine which was used to bond the fill tube and bag. This caused improper sealing which led to media leaking from the fill tube/bag seal only when the bag was placed under stress in a specific packaging configuration. This issue was first identified on june 11th, 2018.
Affected products
OpTmizer CTS T-Cell Expansion SFM
Lot or serial number
1925754
Model or catalog number
A1022103
Companies
- Manufacturer
-
LIFE TECHNOLOGIES CORPORATION ALSO TRADING AS: INVITROGEN
3175 Staley Road
Grand Island
14072
New York
UNITED STATES