Health product recall

Optima NM/CT 640 Unit, Discovery NM/CT (670 CZT, 670 ES), Discovery 670 DR, NM/CT (870 DR, 870 CZT), NM 830 (2019-03-19)

Starting date:
March 19, 2019
Posting date:
April 12, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69540



Last updated: 2019-04-12

Affected Products

  1. Optima NM/CT 640 Unit
  2. Discovery NM/CT 670 CZT
  3. Discovery NM/CT 670 ES
  4. Discovery 670 DR
  5. NM/CT 870 DR
  6. NM/CT 870 CZT
  7. NM 830

Reason

During installation of certain nuclear medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy. Multiple other design redundancies are in place to ensure the detector mounting remains secure. As a result, it is highly improbable that the detector would become unsecure if the screws loosen. In the unlikely event that the screws loosen and the multiple other detector mounting design redundancies fail, the detector could fall during use and lead to potentially life-threatening bodily harm. No detector falls have been reported as a result of this issue.

Affected products

  1. Optima NM/CT 640 Unit

Lot or serial number

ONMX26480

ONMX26502

Model or catalog number

H3100YD

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


B. Discovery NM/CT 670 CZT

Lot or serial number

AD7X38019

Model or catalog number

H3906CU

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


C. Discovery NM/CT 670 ES

Lot or serial number

ESDX35020

Model or catalog number

H3100BK

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


D. Discovery 670 DR

Lot or serial number

DRGX56004

Model or catalog number

H3100AL

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


E. NM/CT 870 DR

Lot or serial number

870Y64034

Model or catalog number

H3100AR

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


F. NM/CT 870 CZT

Lot or serial number

87CX63015

Model or catalog number

H3906CV

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL


G. NM 830

Lot or serial number

830X60022

Model or catalog number

H3910AB

Companies
Manufacturer

GE Medical Systems Israel, Functional Imaging

4 Hayozma Street

Tirat Hacarmel

0030200

ISRAEL