Health product recall

Optima MR360 Main Unit (2014-12-29)

Starting date:
December 29, 2014
Posting date:
January 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43165

Recalled Products

Optima MR360 Main Unit

Reason

Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.

Affected products

Optima MR360 Main Unit

Lot or serial number

403541MR

450670MR2

604985MR1

AM12O9T240

AM12O9T294

Model or catalog number

5312366 (M50002BB)

Companies
Manufacturer

GE Hangwei Medical Systems Co. Ltd.

No. 2 North Yong Chang Street, Beijing Economic, Technological Development Zone,

Beijing

100176

CHINA