ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5 (2015-12-18)
- Starting date:
- December 18, 2015
- Posting date:
- January 18, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56678
Affected Products
A. ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5
Reason
When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
Affected products
A. ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5
Lot or serial number
19311-EB001
Model or catalog number
170.730
Companies
- Manufacturer
-
Nucletron B.V.
Waardgelder 1
Veenedaal
3905 TH
NETHERLANDS