This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Omega Plus Standard Lag Screw

Starting date:
April 17, 2017
Posting date:
May 2, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63152

Reason

Stryker received a complaint regarding a Standard Lag Screw Omega 110mm length. It was reported that the lag screw was not completely cannulated and the guide wire could not go through the cannula of the screw. No harm reported. The investigation revealed that the obstruction is related to a metal burr in the cannula of the screw which was not removed during manufacturing. The non-conformance is limited to 2 batches.

Affected products

Omega Plus Standard Lag Screw

Lot or serial number

V06269

V06279

Model or catalog number

3362-5-105

3362-5-110

Companies

Manufacturer

Stryker GMBH

Bohnackerweg 1,

Selzach, Solothurn

2545