This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Olerup SSP - SSP Combi Trays (2014-05-29)

Starting date:
May 29, 2014
Posting date:
July 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40443

Recalled products

  1. Olerup SSP - SSP Combi Trays 

Reason

Complaint received on kit used for evaluation testing in ring trial (testing with known DNA's in France). A*74:07 allele is not amplified by primer mix 19 as stated in product documentation. This has been corrected in the specificity and interpretation tables. The false negative yields a revised reactivity pattern for the A*74:07 allele that is identical to the A*32:01:14, A*32:03 and A*32:38 pattern.

Affected products

A. Olerup SSP - SSP Combi Trays 

Lot or serial number
  • Lot 64S
Model or catalog number
  • 101.702-24U
Companies
Manufacturer
Olerup SSP AB
Franzengatan 5
Stockholm, Sweden, 11251