Health product recall

OBM - Digital Acquisition Box

Last updated

Summary

Product
OBM - Digital Acquisition Box
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

OBM - Digital Acquisition Box

S/N: 001 through OBM00002H3613

OBM00002

Issue

Natus has become aware of two complaints related to the Olympic Brainz Monitor where the electrode placement labels on the data acquisition box (DAB) were on backwards. This labelling error may lead to a use error where the electrodes are connected incorrectly, causing the left and right channels to be misaligned.

Recall Start Date: May 5, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies

Natus Medical Incorporated Dba Excel-Tech Ltd (Xltek)

2568 Bristol Circle, Oakville, Ontario, Canada, L6H 5S1

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-77455

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe