This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

N/T Protein Standards SL

Starting date:
October 28, 2016
Posting date:
November 15, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61050

Affected Products

Reason

Siemens Healthcare Diagnostics has confirmed that control values may be found out of range (>+/-15%) for N Antiserum to prealbumin using N/T protein control SL/H. Furthermore, controls may be flagged with a turbidity flag (x-flag). We have no indication that other analytes in N/T protein control SL/H are affected by this issue.

Affected products

N/T Protein Standards SL

Lot or serial number
  • 084847
  • 084847C
  • 084847D
  • 084847E
  • 084847F
Model or catalog number
  • OQIP13
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Gmbh

Emil-Von-Behring-Str. 76

Marburg

35041

GERMANY