This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

N/T Protein Controls and Standards (2017-05-31)

Starting date:
May 31, 2017
Posting date:
July 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63800

Affected Products

A. N/T Protein Controls Sl

B. N/T Protein Standards Sl

Reason

Siemens Healthineers has confirmed a negative bias of approximately -20% for results obtained with n antiserum to human ceruloplasmin when calibrating ceruloplasmin with affected N/T protein standard lots. N/T protein controls SL for ceruloplasmin are similarly biased.

Affected products

A. N/T Protein Controls Sl

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

OQIM13

Companies
Manufacturer

Siemens Healthcare Diagnostics Products Gmbh

Emil-Von-Behring-Str. 76

Marburg

35041

GERMANY



B. N/T Protein Standards Sl

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

OQIN13

OQIO13

OQIP13

Companies
Manufacturer

Siemens Healthcare Diagnostics Products Gmbh

Emil-Von-Behring-Str. 76

Marburg

35041