Novum Iq Lvp
Brand(s)
Last updated
Summary
Product
Novum Iq Lvp
Issue
Medical devices - Mechanical defect
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected Products | Lot or serial number | Model or catalog number |
|---|---|---|
| Novum Iq Lvp | All lots. | 40700BAX |
Issue
Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has identified an additional scenario where a rapid double key press will result in unintended behavior on the Novum IQ LVP. This issue will occur if the user rapidly double presses the 'OK' key (<200 milliseconds apart) when accepting a value exceeding the drug library soft limit.
Recall start date: December 29, 2021.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63792