Nipro GlucoPro Insulin Syringes - Recall - For the Public
- Starting date:
- January 29, 2010
- Posting date:
- February 19, 2010
- Type of communication:
- Public Communication
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Audience:
- General Public
- Identification number:
- RA-19000871
This is duplicated text of a letter from Surgo Surgical Supply.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
Public Communication - Health Canada Endorsed Important Safety Information on Nipro GlucoPro Insulin Syringes
January 29, 2010
Subject: Recall of Nipro GlucoPro Insulin Syringes
Surgo Surgical Supply is conducting a recall of all GlucoPro Insulin Syringes. These syringes may have needles that detach from the syringe. If the needle becomes detached from the syringe during use, it can become stuck in the insulin vial, push back into the syringe, or remain in the skin after injection. As a result, use of affected syringes could lead to patient injury or interruption of medication administration.
-
Nipro GlucoPro Insulin Syringes may have needles that detach from the syringe during use.
-
Use of affected syringes could lead to patient injury or interruption of medication administration.
-
Users who have GlucoPro Insulin Syringes with expiration dates on or before 2011-11 (November 1, 2011) should stop using them immediately and return them to the seller for reimbursement or to the clinician who provided them.
| Product Code | Lot # | Expiry date |
|---|---|---|
| JD+01U3008-5C | A08022 | 2011-11 |
| JD+01U3013-5C | A08013 | 2011-06 |
| JD+01U3013-5C | A08017 | 2011-08 |
| JD+01U3108-5C | A08013 | 2011-06 |
| JD+01U3108-5C | A08017 | 2011-08 |
| JD+03U3008-5C | C08022 | 2011-11 |
| JD+03U3013-5C | C08013 | 2011-06 |
| JD+03U3013-5C | C08017 | 2011-08 |
| JD+03U3108-5C | C08013 | 2011-06 |
| JD+03U3108-5C | C08017 | 2011-08 |
| JD+05U3008-5C | B08022 | 2011-11 |
| JD+05U3013-5C | B08013 | 2011-06 |
| JD+05U3013-5C | B08017 | 2011-08 |
| JD+05U3108-5C | B08013 | 2011-06 |
| JD+05U3108-5C | B08017 | 2011-08 |
Action to be taken by the user:
-
Users who have GlucoPro Insulin Syringes with expiration dates on or before 2011-11 (November 1, 2011) should stop using them immediately.
-
All affected syringes should be returned to the seller for reimbursement or to the clinician who provided them.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-market adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of needle detachment or other serious or unexpected adverse incidents for medical devices in patients using Nipro GlucoPro Syringes should be reported to Surgo Surgical Supply or Health Canada at the following addresses:
Surgo Surgical Supply
183 Simcoe Ave
Keswick ON L4P 3S7
Telephone: 1-800-663-0395
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: MHPD_DPSC@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
original signed by
Hugo Jeffrey
Surgo Surgical Supply