Health product recall

Nimbus 4 & Nimbus Professional mattresses (2021-04-13)

Brand(s)
Last updated

Summary

Product
Nimbus 4 & Nimbus Professional mattresses (2021-04-13)
Issue
Medical devices - Manufacturing defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
NIMBUS 4 MATTRESS More than 10 numbers, contact manufacturer. 651001DAR
NIMBUS PROFESSIONAL MATTRESS More than 10 numbers, contact manufacturer. 651001DAR
NIMBUS PROFESSIONAL PREMIUM MATTRESS More than 10 numbers, contact manufacturer. 651001P

Issue

Customer complaint on over-inflation (abnormal inflation issue) occurring during the product installation, preparation to initiate therapy or product use. In none of those events, any injury or health consequences were reported. The products in the market have been recognized as potentially affected with the risk of abnormal inflation which might result in the mattress surface raised above its specification. The investigation revealed that the failure is caused by damage of inner mattress tubes.

Recall start date: Apr 13, 2021

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

ARJO CANADA INC.

90 MATHESON BLVD. WEST, SUITE 350, MISSISSAUGA, ON

CA, L5R 3R3

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-62377

Get notified

Receive emails about new and updated recall and safety alerts.