Health product recall

neoBLUE mini LED Phototherapy System, neoBLUE LED Therapy Device, neoBLUE Cozy, Blanket, and Compact LED Phototherapy System (2018-04-08)

Starting date:
April 8, 2018
Posting date:
May 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66626

Affected products

A. NEOBLUE MINI LED PHOTOTHERAPY SYSTEM

B. NEOBLUE COZY LED PHOTOTHERAPY SYSTEM

C. NEOBLUE LED THERAPY DEVICE

D. NEOBLUE BLANKET LED PHOTOTHERAPY SYSTEM

E. NEOBLUE COMPACT LED PHOTOTHERAPY SYSTEM

Reason

Natus is updating the Instructions For Use for all of the neoBLUE LED Phototherapy devices to include the information that is important to measure the intensity with a radiometer specifically designed to measure the narrow wavelength spectrum of blue LED lights. Radiometers designed to measure the broadband spectrum found in fluorescent or halogen lights may result in inaccurate intensity measurements for neoBLUE LED products.

Affected products

A. NEOBLUE MINI LED PHOTOTHERAPY SYSTEM

Lot or serial number

Not applicable

Model or catalog number

010101

Companies
Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES



B. NEOBLUE COZY LED PHOTOTHERAPY SYSTEM

Lot or serial number

Not applicable

Model or catalog number

010121

Companies
Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES



C. NEOBLUE LED THERAPY DEVICE

Lot or serial number

Not applicable

Model or catalog number

010066

010068

010069

010070

Companies
Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES



D. NEOBLUE BLANKET LED PHOTOTHERAPY SYSTEM

Lot or serial number

Not applicable

Model or catalog number

006244

006898

 

Companies
Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES



E. NEOBLUE COMPACT LED PHOTOTHERAPY SYSTEM

Lot or serial number

Not applicable

Model or catalog number

019001

019011

Companies
Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES