NDI Passive Spheres™
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
NDI Passive Spheres™ |
More than 10 numbers, contact manufacturer. |
8801071 |
NDI Passive Spheres™ |
More than 10 numbers, contact manufacturer. |
8801074 |
NDI Passive Spheres™ |
More than 10 numbers, contact manufacturer. |
8801075 |
Issue
The sphere manufacturer has identified variation in the reflective material used on a limited number of sterile passive sphere lots. The reflective material is the outer coating of the sphere that reflects infrared light from the optical tracking cameras, enabling detection and tracking of instruments during navigation procedures. Medtronic has determined that this variability may affect the ability of the navigation system to detect and track the spheres. Potential impacts may include instrument verification failures, intermittent or lost tracking, red status tracking or navigation accuracy.
Additional information
Details
Northern Digital Inc.
103 Randall Drive, Waterloo, Ontario, Canada, N2V 1C5