This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Navios Color Lasers (2013-05-30)

Starting date:
May 30, 2013
Posting date:
October 16, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36053

Recalled Products

A. Navios 6 Colors / 2 Lasers 

B. Navios 8 Colours / 2 Lasers

C. Navios 10 Colours / 3 Lasers

Reason

Beckman Coulter has identified that the fluorescence photo-multiplier tube (PMT) gain settings displayed in the cytometer control dialog box may not reflect the actual PMTgain settings on the Navios flow cytometer. If a change to the fluorescence gain is made at any time to any protocol and saved (or not saved), all susequently loaded protocols will apply the previously set gain parameters regardless of what the cytometer control dialog is showing. Therefore, any subsequent protocol may not reflect the actual system gain in the cytometer control dialog box.

Affected products

A. Navios 6 Colors / 2 Lasers

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • A52101
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



B. Navios 8 Colours / 2 Lasers

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • A52102
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



C. Navios 10 Colours / 3 Lasers

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • A52103
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES