This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Navios 8 and 10 Colour Lasers (2013-08-28)
- Starting date:
- August 28, 2013
- Posting date:
- September 19, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-35751
Recalled Products
A. Navios 8 Colors / 2 Lasers
B. Navios 10 Colors / 3 Lasers
Reason
Beckman Coulter has determined that when the maintenance log for December is printed or saved, entries for the subsequent eleven (11) months are lost and cannot be retrieved unless the database was backed-up prior to performing the print or save actions. For example, if December 2012 is printed or saved, any maintenance log entries for January-November 2013 will be lost. This issue can result in potentially incomplete maintenance log entries. No other report generator functions or patient results are affected by this issue.
Affected products
A. Navios 8 Colors / 2 Lasers
Lot or serial number
All lots
Model or catalog number
- A52102
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Navios 10 Colors / 3 Lasers
Lot or serial number
All lots
Model or catalog number
- A52103
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES