Health product recall

Namic Stopcock/Manifold (2021-08-18)



Starting date:
August 18, 2021
Posting date:
September 10, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76395


Last updated:

Affected Products

  1. Namic Stopcock
  2. Namic Manifold

Reason

During the testing process, a breach was confirmed in the Sterile barrier via Micro-hole testing, resulting in

the possibility of non-sterile product conditions. The microscopic defects in the Mylar are

6420 70055003 h965700550031

sc 3w hp fm off nb -

pg

0000007025, 0000012635, 0000016518,

5653288, 5658901, 5662967

6420 70055008 h965700550081

sc 3w hp ra/lt off nb

- pg

0000007081, 0000014807, 5653287,

5656299

6420 70055009 h965700550091

sc 3w hp ra off nb -

pg

0000003281, 0000003284, 0000007024,

0000007600, 0000007601, 0000009626,

0000009627, 0000010746, 0000012300,

0000012301, 0000012302, 0000014982,

0000015022, 0000016270, 0000016520,

5653286, 5657315, 5658895, 5658897,

5659900, 5659903, 5659905

6420 70055017 h965700550171

sc 1w hp fm tee nb -

pg

0000007147, 0000010747, 0000015023,

56619973

Medline Canada

5150 Spectrum Way, Suite 300

Mississauga, ON I4W 5G2



Too small to be detected visually with the naked eye, therefore we cannot depend on users to determine whether a unit can be used.

Affected products

  1. Namic Stopcock

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • 70015002
  • 70015012
  • 70015015
  • 70035008
  • 70055003
  • 70055009
  • 70055017
Companies
Manufacturer

Medline Industries Inc.

Three Lakes Drive

Northfield

60093

Illinois

UNITED STATES



B. Namic Manifold

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • 70037200
  • 70038202
  • 70038303
  • 70039303
Companies
Manufacturer

Medline Industries Inc.

Three Lakes Drive

Northfield

60093

Illinois

UNITED STATES