This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

N Latex B2 Microglobulin (2014-10-01)

Starting date:
October 1, 2014
Posting date:
October 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41747

Recalled Products

N Latex B2 Microglobulin

Reason

Siemens Healthcare Diagnostics has received a small number of complaints about calibration failure with N Latex B2 Microglobulin. In rare cases the signal generated at the highest calibration point can exceed the analog-digital converter capacity of an affected BN system. This effect does not allow the establishment of a calibration curve due an invalid calibration point measurement.

Affected products

N Latex B2 Microglobulin

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

OQWU15

Companies
Manufacturer

Siemens Healthcare Diagnostics Products GmbH

Emil-Von-Behring-Str. 76

Marburg

35041

GERMANY